The quick answer is yes and no. Yes, much of what they recommend is valid
for all measurement methods. No, they were written
for open-chamber instruments and the AquaFlux is different. If you want
the long answer, read on.
These Guidelines [1, 2] pre-date the invention of the AquaFlux.
Guidelines [1] consider only the Evaporimeter Model EP1 manufactured
by Servomed, whereas Guidelines [2] include newer open-chamber
instruments.
Both publications discuss (a) person-linked variables, (b) environmental
variables and (c) instrumental variables in some depth before presenting
their recommendations for best practice. Most of the considerations
of sections (a) and (b) are valid for all methods of measurement. The
exceptions are section (a) of [2], where the discussion
of skin surface temperature includes consideration of how this may
affect the measurement probe, and section (b) of both reports, which
include discussions of air circulation and how this may cause
fluctuations in (open-chamber) measurements. Neither of these are concerns
for the AquaFlux. Section (c) and elsewhere is instrument-specific
and not applicable to the AquaFlux.
The table below presents a side-by side comparison of the recommendations
of these guidelines with what we recommend for
the AquaFlux.
|